Concepts and features
R&S
®
ZNA
275
User Manual 1178.6462.02 ─ 20
IF BW
(MHz)
Time resolution
(ns)
Simultaneously measured wave quantities
2
4
8
16
0.1
552
✓
✓
✓
✓
0.15
368
✓
✓
✓
with B7
1
48
✓
✓
✓
with B7
1.5
32
✓
✓
✓
with B7
2
24
✓
✓
with B7
10
16
✓
✓
with B7
15
16
✓
✓
with B7
20
8
✓
with B7
30
8
✓
with B7
The number of wave quantities that are required to measure a certain quantity,
depends on the setup and the calibration. E.g., a calibrated gain measurement with
"One Path Two Ports" calibration requires the simultaneous measurement of three
wave quantities.
To retrofit the memory extension board, the R&S
ZNA must be sent to Rohde &
Schwarz Service.
4.7.24
LO Out
R&S
ZNA-B8
Hardware option R&S
port.
This port is particularly useful for driving millimeter wave converters. If you use a power
splitter to connect the LO Out of the R&S
ZNA to the LO In ports of the converters,
then it is possible to measure n converter ports with an n-port R&S
ZNA, without the
need for an external generator.
If the R&S
ZNA is equipped with a second internal LO generator, then the second inter-
nal LO generator is connected to the LO Out.