MIMPedance
"Imp Mag" for reflection measurement traces
"Imp Mag series" for transmission measurement
traces
MPIMpedance
"Imp Mag" for reflection measurement traces
"Imp Mag parallel" for transmission measurement
traces
MADMittance
"Adm Mag" for reflection measurement traces
"Adm Mag series" for transmission measurement
traces
MPADmittance
"Adm Mag" for reflection measurement traces
"Adm Mag parallel" for transmission measurement
traces
INDex
Index
NOISe
Noise
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPECtrum
For marker <Mk> of channel <Chn>'s active trace, this command does the following:
●
Create a channel
●
Set its center to marker <Mk>'s position
●
Set its span to 10 times channel <Chn>'s BW
●
Switch to spectrum mode
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number
Usage:
Event
Options:
R&S
ZNA-K1
Manual operation:
See
CALCulate<Chn>:MARKer:MPEak:EXCursion
<TargetSearchVal>
Defines a minimum excursion value for multiple peak searches.
Use
CALCulate<Chn>:MARKer:MPEak:EXCursion:STATe
to activate it.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Value>
Minimum peak excursion
The unit is derived from the active trace format and cannot be
modified. Changing the trace format resets the excursion to a
format-specific default value.