GUI reference
R&S
®
ZNA
415
User Manual 1178.6462.02 ─ 20
Ref. Range
This button is visible if either:
●
statistics are calculated and "Range" is used as
●
The current channel is in spectrum analysis mode
It opens the "Reference Range" dialog that allows you to configure the sweep range
whose average value shall be used as the reference ("small signal value") for the com-
pression point calculation or as the reference power for the
calculation.
Remote command:
Compression point statistics:
CALCulate<Chn>:STATistics:NLINear:COMP:RDOMain:USER
CALCulate<Chn>:STATistics:NLINear:COMP:RDOMain:USER:STARt
CALCulate<Chn>:STATistics:NLINear:COMP:RDOMain:USER:STOP
CALCulate<Chn>:STATistics:NLINear:COMP:RDOMain:USER:SHOW
Power ratio statistics (spectrum analysis) :
CALCulate<Chn>:STATistics:RDOMain:USER
CALCulate<Chn>:STATistics:RDOMain:USER:STARt
CALCulate<Chn>:STATistics:RDOMain:USER:STOP
CALCulate<Chn>:STATistics:RDOMain:USER:SHOW
Defined Value
This setting is only visible if "Defined Value" is used as
. It manually
defines the reference value ("small signal value") for the compression point calculation.
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:RLEVel
CALCulate<Chn>:STATistics:NLINear:COMP:RPHase
5.5.8.2
Evaluation Range dialog
The "Evaluation Range" dialog defines the range for various trace evaluation opera-
tions (trace statistics, offset, loss and fixture compensation calculations). The evalua-
tion range is a continuous interval of the sweep variable.
Traces softtool