GUI reference
R&S
®
ZNA
610
User Manual 1178.6462.02 ─ 20
The measurement (=sampling) time can only be set via remote command. It is not
channel-specific and applies to RFFE and GPIO measurements.
Remote command:
CONTrol<Ch>:GPIO:SENSe:TRIGger
CONTrol<Ch>:GPIO:TIME
Voltage, Current
Results of the voltage/current measurements on the related GPIO pins.
Remote command:
CONTrol:GPIO<Port>:SENSe:VOLTage?
CONTrol:GPIO<Port>:SENSe:CURRent?
CONTrol:GPIO:SENSe:SUMCurrent?
5.12.1.6
Sweep Sequencer
Defines the RFFE commands to be executed and the GPIO voltages to be applied
before each sweep (segment).
●
Make sure that the required RFFE bus interfaces and GPIO ports are enabled by
setting the respective "Seq." flags in the "RFFE Config" dialog (see
interface settings and command execution"
on page 604 and
).
●
The sequencer settings are only valid for the active channel. In a multiple channel
measurement, you have to configure each channel separately.
●
On a R&S
ZN-B15/Z15
var. 03
, the sweep sequencer overwrites the currently
applied voltages and sets the related shunt resistances to the minimum possible
value (see
"RFFE interface voltage and current measurements"
on page 606 and
"GPIO voltage and current measurements"
Proceed as follows:
1. For segmented sweeps, select the related sweep segment (see
Channel Config softtool