GUI reference
R&S
®
ZNA
611
User Manual 1178.6462.02 ─ 20
2. Use the "Add" or "Insert" button to add a new step to the command/switch
sequence - either at the end of the existing sequence or above the selected step,
respectively.
3. Make the appropriate settings in the sweep sequencer table:
a) Define the RFFE commands ("RFFE" columns).
b) Set the GPIO voltages ("GPIO" columns).
c) Specify the delays between subsequent commands/switches and before the
sweep or segment start ("Wait" column).
4. Use "Delete" to remove the selected item or "Delete All" to clear or the command/
switch sequence.
5. Select "OK" to apply the settings. To discard the changes, select "Cancel" .
Range
For segmented sweeps, a command/switch sequence can be defined for each sweep
segment. For unsegmented sweeps, only a single "Range" is available.
Remote command:
Unsegmented sweeps
CONTrol<Ch>:SEQuence:COUNt?
CONTrol<Ch>:SEQuence:CLEar:ALL
Segmented sweeps:
CONTrol<Ch>:SEGMent<Nr>:SEQuence:COUNt?
CONTrol<Ch>:SEGMent<Nr>:SEQuence:CLEar:ALL
RFFE columns (sweep sequencer table)
The cells in the "RFFE" columns define the sequence of commands to be sent over the
respective RFFE interface. The command definition dialog (opened from the ellipsis
buttons) is described in
"Basic RFFE interface settings and command execution"
Remote command:
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe]
Unsegmented sweeps
CONTrol<Ch>:SEQuence<Nr>:RFFE<Bus>:COMMand:DATA
Segmented sweeps:
CONTrol<Ch>:SEGMent<Nr>:RFFE<Bus>:COMMand:DATA
GPIO columns (sweep sequencer table)
The cells in the "GPIO" columns define the sequence of voltages to be applied to the
respective GPIO ports.
Remote command:
CONTrol<Ch>:GPIO<Port>[:STATe]
Unsegmented sweeps
CONTrol<Ch>:SEQuence<Nr>:GPIO<Port>:VOLTage
Segmented sweeps:
CONTrol<Ch>:SEGMent<Nr>:GPIO<Port>:VOLTage
Channel Config softtool