Command reference
R&S
®
ZNA
1453
User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored; the generator is
selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer:CORRection:IMODulation:METHod
<FastOrNot>
Enables or disables the fast interpolation method for intermodulation calibration.
Suffix:
<Ch>
.
Channel number
Parameters:
<FastOrNot>
ON (1)
Fast method enabled
OFF (0)
Fast method disabled
*RST:
ON (1)
Options:
R&S
ZNA-K4
Manual operation:
See
"IMD Cal Method>Fast Cal Interp. Meas"
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
Starts the source calibration for the lower tone (1
st
power calibration step for intermo-
dulation measurements), stores and applies the calibration data. The external power
meter used is selected via
SOURce<Ch>:POWer<PhyPt>:CORRection:PMETer:ID
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number This suffix is ignored; the analyzer calibrates the
port that is selected as a source port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
).
SCPI command reference