GUI reference
R&S
®
ZNA
637
User Manual 1178.6462.02 ─ 20
5.13.2.1
Controls on the Pulse Mod tab
Pulse Generator
Enables or disables pulse generation.
Remote command:
[SENSe<Ch>:]PULSe:GENerator<Id>[:STATe]
Set all Traces to Matching Detector
Tells the firmware to select a suitable detector for all traces in the current channel. For
background information, see
Chapter 4.3.5.3, "Detector settings"
VSE Settings
The controls in the "VSE Settings" part of the "Pulse Mod" tab allow you to set up, con-
figure and run
Chapter 4.7.6.1, "Live pulse analysis with R&S
Configure VSE ← VSE Settings
VSE Connection Configuration dialog
that allows you to establish a connec-
tion to a local or remote R&S
VSE instance.
VSE Mode ← VSE Settings
Determines where the acquired measurement data are processed.
"ZNA Only"
(default)
The selected
ZNA; no data are
pushed to the configured R&S
VSE instance.
"VSE Only"
The R&S
ZNA does not calculated the selected trace. Instead, it pre-
pares I/Q data for the measured b-wave and pushes them to the
R&S
VSE for further analysis.
Mode softtool