Command reference
R&S
®
ZNA
1241
User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number
Usage:
Query only
Options:
R&S
ZNA-K4
Manual operation:
See
"Harmonic Orders to Calibrate"
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD
<Order>
Adds a harmonic order to the list of harmonic orders at which source and receiver cali-
brations shall be performed for harmonics measurements in channel <Ch>.
Use
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
to get this
list, and
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
to
remove a harmonic order from this list.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
) to calibrate.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Order>
Order number
Usage:
Setting only
Options:
R&S
ZNA-K4
Manual operation:
See
"Harmonic Orders to Calibrate"
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
Removes a harmonic order from the list of harmonic orders at which source and
receiver calibrations shall be performed for harmonics measurements in channel <Ch>.
Use
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
to get this
list, and
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD
to add a
harmonic order to it.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
) to calibrate.
Suffix:
<Ch>
.
Channel number
Usage:
Event
Options:
R&S
ZNA-K4
Manual operation:
See
"Harmonic Orders to Calibrate"
SCPI command reference