Command reference
R&S
®
ZNA
1227
User Manual 1178.6462.02 ─ 20
Parameters:
<AverageFactor>
Sweep average factor
Range:
1 to 1000
*RST:
10
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:MODE
<Mode>
Suffix:
<Ch>
.
Parameters:
<Mode>
AUTO | FLATten | REDuce | MOVing
AUTO
Automatic selection between REDuce and FLATten mode,
depending on the trace format.
FLATten
Cumulative moving averages of the (linear) magnitude and
phase values, provides the most effective noise suppression for
the "dB Mag", "Phase", "Unwr Phase", and "Lin Mag" formats.
REDuce
Cumulative moving averages of the real and imaginary parts of
each measurement result, provides the most effective noise sup-
pression for the "Real" and "Imag" formats and for complex trace
formats.
MOVing
Simple moving averages of the real and imaginary parts of each
measurement result; similar to REDuce, but with finite history.
Manual operation:
See
7.3.14.2
[SENSe:]BANDwidth...
The
[SENSe:]BANDwidth...
commands set the bandwidth of the IF filter (measure-
ment bandwidth).The forms
BANDwidth
and
BWIDth
are equivalent.
[SENSe<Ch>:]BANDwidth[:RESolution]
..........................................................................1228
[SENSe<Ch>:]BWIDth[:RESolution]
...............................................................................1228
[SENSe<Ch>:]BANDwidth[:RESolution]:DREDuction
.......................................................1228
[SENSe<Ch>:]BWIDth[:RESolution]:DREDuction
............................................................ 1228
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
.............................................................. 1228
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
....................................................................1228
SCPI command reference