GUI reference
R&S
®
ZNA
537
User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
IMD Cal Method>Fast Cal Interp. Meas
This checkbox is enabled, if the channel is configured for an intermodulation (IMD)
measurement.
To avoid interpolation of calibration data, which causes problems during the measure-
ment, it is preferable to increase the number of sweep points during the calibration.
However, the number of sweep points that are required to avoid interpolation can be
large, and hence the calibration can take some time to complete.
Keep "Fast Cal Interp. Meas" checked to stick to the faster calibration method, that
does not increase the number of sweep points and uses interpolation if necessary.
Remote command:
SOURce<Ch>:POWer:CORRection:IMODulation:METHod
Calibration Mixer Settings
Defines the properties of the calibration mixer and the related calibration. The table is
enabled, if the channel is configured for a mixer measurement.
Generic Device Setting for Calibration Mixer
Allows you to execute a command on a previously configured
the calibration mixer step of a calibration.
This setting can be used, for example, to insert the calibration mixer into the signal
path using an external switching unit such as the R&S
OSP.
Remote command:
Same Sweep Setup for all Standards
Same Sweep Setup for All Standards
logic for the current channel (instead
globally).
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CHANnels:CSETup
Cal softtool