Command reference
R&S
®
ZNA
1378
User Manual 1178.6462.02 ─ 20
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Usage:
Event
Manual operation:
See
"Add / Insert / Delete / Delete All"
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
<ResBandwidth>
Defines the IF bandwidth of the analyzer (measurement bandwidth) in sweep segment
no.
<Seg>
. Values between 1
Hz and 1.5
MHz can be set. Option R&S
ZNA-K17 ena-
bles IF bandwidths up to 10
Chapter 4.7.8, "Increased IF bandwidth
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The analyzer rounds up any
entered value between these steps and rounds down values exceeding the maximum
bandwidth.
At the same time, the command activates separate bandwidth setting in all sweep seg-
ments (
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
ON
).
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<ResBandwidth>
IF bandwidth
Range:
See above
Increment: 1-1.5-2-3-5-7 steps
*RST:
10 kHz
Default unit: Hz
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
<Boolean>
Selects common or independent "Meas. Bandwidth" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
SCPI command reference