background image

 

 

 

GN128-Explosion-Proof Piston Rev. 0 

Orange Research Inc. 

Page 1 of 2

 

INSTRUCTIONS 

Series 1204 

DIFFERENTIAL PRESSURE INSTRUMENTS 
Piston Sensor 

– Explosion Proof 

 

Your new Orange Research Differential Pressure Instrument is a 
rugged instrument featuring simplicity of design to provide 
dependable and efficient service.  Because it is an instrument it 
should be handled with care.  Read all instructions carefully 
before attempting to install the instrument. 

CAUTION

: Do not exceed maximum operating pressure 

listed on instrument label.  Use only fluids compatible with 
wetted parts, including seals. 

HOW IT WORKS 

The instrument operates on the difference between two 
pressures (delta-P).  The sensing element is a spring biased 
piston which moves linearly in proportion to the difference 
between two basic pressures.  A magnet on the 

HI

 pressure side 

of the piston assembly moves with the piston and rotates a 
follower magnet located adjacent to the pressure cavity.  The 
gauge pointer is located at the end of the rotary magnet shaft and 
rotates with the magnet to provide gauge readings proportional to 
differential pressure variations.  This is accomplished by the 
coupling forces between two adjacent magnets through a solid 
wall. 

SWITCHES:

 

A secondary magnet, located in the extension tube 

also moves with the pressure sensor. Reed switches, mounted 
externally on the pressure extension tube are actuated when the 
field of the secondary magnet interacts at a preset point with the 
reed switch armatures. 

NOTE:

 This instrument will provide 

2% accuracy full scale. 

INSTALLATION 

Check instrument and identify the 

HI

 and 

LO

 markings.  

HI

 

identifies the high pressure port; 

LO

 the low pressure port.  If 

instrument is installed backwards, it will neither operate nor be 
damaged. The instrument can be line mounted, pipe mounted or 
panel mounted. 
 

Installation Notes: 

 

These piston sensor instruments are design for liquid 
service only…use in gas systems will result in erroneous 
readings due to excessive piston bypass 

 

Do not mount directly to steel surfaces…this could affect 
the magnetic sensors…mounting 1 inch from ferrous 
metals is recommended 

 

If there is particulate in the line, insert a screen (approx. 
100 micron) to block the particulate from becoming 
lodged in the instrument 

 

Do not install on systems for highly viscous fluids or 
where the fluid may become hardened over time 

 

Do not install on systems with iron in the fluid…the iron 
can become attached to the magnetic sensors 

 

Install the unit above the process lines to prevent 
particulate from settling in the instrument 

 

Before installation, refer to our Specification Sheets (on 
our website) to identify instrument component materials 
& double check for chemical compatibility 

 
 

 
 

 

 

 

 

 

MAINTENANCE 

Other than replacing a broken lens, there is only one area where 
this instrument may need attention. Erratic pointer or switch 
action may indicate that cleaning is required. For cleaning: 
1.  Remove the unit from service. 
2. 

Remove end cap with a 1” spanner wrench. 

3.  Remove the range spring and the piston/magnet assembly. 

BE CAREFUL NOT TO LOSE STACKING SPACERS, IF 
ANY. 

4.  Clean parts 

in a solvent solution after removing the “O”-ring 

seals from the end caps since some solvent will attack the 
seal material. 

5. 

It is good practice to replace the “O”-ring seals while the 
instrument is dismantled. DO NOT ATTEMPT TO REMOVE 
THE TEFLON PISTON SEAL FROM THE PISTON. If piston 
seal is damaged order a new piston/magnet assembly. 

To reassemble: 

1.  Replace the piston assembly. 

CAUTION:

 Do not bend the 

stainless steel wire which connects the secondary magnet to 
the piston assembly. 

2.  Make certain that the spacers are seated in the piston spring 

pocket and reinsert the spring. 

3.  Reassemble the end cap and tighten. 

NOTE:

 Be certain that 

the head of the spring is seated in the end cap spring 
pocket. The unit is now ready for service. 

LENS REPLACEMENT:

 To remove a threaded retaining ring, 

unscrew the retaining ring, or to remove the snap ring, pry out the 
ring with a small screwdriver.  Remove the glass chips, insert 
new lens and reinstall the snap ring and threaded ring. For snap 
rings locate the ring joint at the bottom of the gauge. 

POINTER REPLACEMENT:

  (probably damaged when lens was 

broken). Remove Threaded ring or snap ring as previously 
described and clean out glass chips. Remove old pointer with 
pointer puller or two small screwdrivers opposite each other 
under pointer hub. Pry off evenly being careful not to bend the 
pointer shaft. Install new pointer dead on zero. Re-install lens, as 
described under Lens Replacement.

Reviews: