Command reference
R&S
®
ZNA
925
User Manual 1178.6462.02 ─ 20
Parameters:
<CenterTime>
Center time of the time gate
Range:
-99.8999999 s to +99.8999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON; :CALC:FILT:TIME:
STAT ON
Reset the instrument and enable the time domain representation
and the time gate.
CALC:FILT:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
<SidebandSupp>
Sets the sideband suppression for the Dolph-Chebyshev time gate. The command is
only available if a Dolph-Chebyshev time gate is active (
DCHebyshev
).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SidebandSupp>
Sideband suppression
Range:
10 dB to 120 dB
Increment: 10 dB
*RST:
32 dB
Default unit: dB
Example:
*RST; :CALC:FILT:TIME:WIND DCH
Reset the instrument and select a Dolph-Chebyshev time gate
for filtering the data in the frequency domain.
CALC:FILT:TIME:DCH 25
Set the sideband suppression to 25 dB.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
<TimeGate>
Selects the time gate to be applied to the time domain transform.
Tip:
Use the generalized command
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
if you wish to select a Dolph-Chebychev time gate.
SCPI command reference