Command reference
R&S
®
ZNA
1058
User Manual 1178.6462.02 ─ 20
Options:
R&S
ZNA-K2
Manual operation:
See
"Start Distance / Stop Distance"
CALCulate<Chn>:TRANsform:DTFault:STATe
<DtfState>
If channel <Chn> is performing an unsegmented frequency sweep, this command
allows to transform its active trace into a Distance to Fault (DtF) trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DtfState>
OFF|ON – DtF disabled|enabled
*RST:
OFF
Example:
:CALCulate1:TRANsform:DTFault:STATE ON
turns the DtF representation of the active trace of channel 1 ON.
Options:
R&S
ZNA-K2
Manual operation:
See
7.3.1.25
CALCulate:TRANsform:VNETworks...
The
CALCulate:TRANsform:VNETworks...
commands define and activate the cir-
cuit models for the different de-/embedding types. See
The circuit models are addressed by predefined enum constants. Except for the FIM-
Port model, that consists of a single snp file, the models and constants are different for
the different de-/embedding types.
Table 7-6: Circuit models for single ended port embedding/deembedding
Parameter
Circuit model
Pictogram
FIMPort
File import, generic 2-port (no circuit model)
CSL
Serial C, shunt L
LSC
Serial L, shunt C
SCPI command reference