GUI reference
R&S
®
ZNA
762
User Manual 1178.6462.02 ─ 20
Background information
Refer to
Chapter 4.7.2.7, "Distance-to-fault measurements"
5.17.3.1
Controls on the Distance to Fault tab
The controls on the "Distance to Fault" tab allow you to enable and configure a stan-
dard Distance to Fault (DtF) measurement.
A standard DtF measurement is prepared in the order from top to bottom:
1. Enable DtF measurement, see
2. Configure the distance window, see
"Start Distance / Stop Distance"
3. Adjust the number of sweep points (and, if necessary, the frequency span), see
4. Select (or define and select) a suitable cable type, see
5. Perform a full one-port calibration at physical port 2.
You should now be able to locate the faults (peaks) by examining the trace.
You can also let the firmware generate a list of faults by enabling
and defining a suitable
detected faults.
Distance to Fault
Activates/deactivates Distance to Fault representation for the active trace.
Note
that "Distance to Fault" can only be enabled, if the active channel is configured to
Applic softtool