Command reference
R&S
®
ZNA
1318
User Manual 1178.6462.02 ─ 20
Example:
*RST; CORR:POW:ACQ BWAVe,1,PORT,2
Perform a receiver power calibration of the wave b1 using port 2
as a source port, assuming that the source power a
2
is correct.
A through connection from port 2 to port 1 is needed.
CORR:POW?
Check whether the calibration is applied (the response is
1
).
CALC:PAR:MEAS 'TRC1', 'B1D2'
Select b
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no.
<Mk>
in the center of the sweep range and
query the measurement value. The calibrated power of the
received wave b
1
is approx. equal to the default source power
value.
See also example for
Manual operation:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
<Wave>[, <CalPort>[,
<SourceType>[, <SourcePort>[, <CalOnlyPortFreq>]]]]
Selects the wave quantity and the source for the receiver power calibration, starts the
calibration sweep, and applies the receiver power correction.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
This suffix is ignored because the port number is specified in the
parameter list.
Setting parameters:
<Wave>
AWAVe | BWAVe | B1 | B2 | B3 | B4
AWAVe
–
calibration of reference waves a
1
, a
2
, .... In manual
control, the reference receiver calibration is included in the
source power calibration of each port.
BWAVe
–
calibration of received wave b
1
, b
2
, .... The port num-
ber
<CalPort>
, the used
<SourceType>
, and the
<SourcePort>
number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ...
–
direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No
additional parameters must be specified. The source for B1 is
port 2 and vice versa.
<CalPort>
Calibrated port number
Range:
1 to port number of the analyzer
SCPI command reference