Command reference
R&S
®
ZNA
1240
User Manual 1178.6462.02 ─ 20
[SENSe:]CONVerter<Port>:RFPort?
Returns the VNA source port that drives converter <Port>.
Always returns <Port>.
Suffix:
<Port>
.
Converter number
Usage:
Query only
Options:
R&S
ZNA-K8
Manual operation:
See
[SENSe:]CONVerter<Port>:RFSLope
<Slope>
Suffix:
<Port>
.
Parameters:
<Slope>
Default unit: DB/GHZ
Manual operation:
See
[SENSe:]CONVerter<Port>:STATe
<Boolean>
Activates or deactivates converter <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
Parameters:
<Boolean>
Options:
R&S
ZNA-K8
Manual operation:
See
7.3.14.5
SENSe<Ch>:CORRection:ADVanced
Remote control commands for advanced calibration features.
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
Returns the list of harmonic orders at which the source and receiver power calibration
shall be performed for harmonics measurements in channel <Ch>.
Use
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD
/
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
to add/remove
harmonic orders to/from this list.
Use
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire]
and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
) to calibrate.
SCPI command reference