GUI reference
R&S
®
ZNA
829
User Manual 1178.6462.02 ─ 20
IF Frequency
Defines the intermediate frequency at the related port. Defaults to the
of the
If you specify an IF below 28
MHz, then during spectrum measurements the firmware
automatically adjusts the IF to 31.25
MHz.
Remote command:
[SENSe:]CONVerter<Port>:IFRequency
IF Input
Selects the IF input connectors to be used.
"Dir. Access"
Chapter 4.7.26, "Direct generator/receiver access"
"Rear IF"
Chapter 4.7.29, "Direct IF access"
Remote command:
[SENSe:]CONVerter<Port>:IFPort
Active
Activates/deactivates the converter configuration for the related port.
Remote command:
Source
Displays the related source port on the VNA (read-only).
Remote command:
[SENSe:]CONVerter<Port>:RFPort?
Local
Defines the source for the LO In port of the converter.
In general, if two or more converters are used in a measurement, it is recommended to
distribute a single LO signal to all converters using a splitter. Select "Splitter 1" or
"Splitter 2" in this case and use the intended LO source as source of
"None"
No LO source assigned. Results in an invalid converter configuration
that cannot be activated.
"Port <i>"
Use a VNA port that is not used as converter port as the LO source.
"Gen <i>"
Use an external generator as LO source.
"Splitter <i>"
Distribute the LO source via a splitter. "Splitter 2" is only useful, if 2
independent 2-port converter measurements shall be performed.
"LO Out"
at the rear panel as LO source.
If compatible with the overall converter setup, the source providing the LO IN signal to
a frequency converter is automatically configured as permanent (
).
Remote command:
[SENSe:]CONVerter<Port>:LOPort
Setup softtool