calibration. You can also switch to another channel and configure/start this
channel's calibration (see
Chapter 4.2.2.3, "Multi-channel setup dialog"
●
"Calibrate All" section
The main purpose of the buttons in this section is to perform complex multi-channel
calibrations in an efficient way.
–
"Quick Start Calibrate All..." immediately starts the calibration, using a previ-
ously configured subset of the available calibration setups, or all available cali-
bration setups.
Opens the
that guides you through the calibration data acqui-
sition.
–
"Configure/Start Calibrate All..." opens the
that allows you
to configure calibration setups for an arbitrary set of channels and to enable/
disable a subset of the configured calibration setups. Finally you can acquire
calibration data for the enabled calibration setups in an efficient way.
●
"Power Cal Only" section
The "Scalar Power Cal" button opens the
perform "Ref. Receiver", "Meas. Receiver" and scalar source power ("Source Flat-
ness") calibrations for the current channel.
Guided Calibration...
Use "Start... (Guided)" to open the
that allows you to config-
ure and run a set of calibrations efficiently.
Cal Validation ...
that allows you to set up and run a
.
5.11.1.3
Calibration wizard
Guides you through the acquisition of calibration data (system error correction and
power correction).
Access:
●
Channel – [Cal] > "Start Cal" > "Cal Active Channel" – "Quick Start Calibration"
●
> "Start Cal"
●
Channel – [Cal] > "Start Cal" > "Calibrate All" – "Quick Start Calibrate All"
●
> "Start Cal"
If accessed via "Quick Start Calibration" or the calibration setup dialog, a single chan-
nel is calibrated. If accessed via "Quick Start Calibrate All" or the "Calibrate All" dialog,
possibly multiple channels are calibrated. Depending on the calibration setup, the cali-
bration can also involve several power calibration steps.
Cal softtool