GUI reference
R&S
®
ZNA
529
User Manual 1178.6462.02 ─ 20
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
[SENSe<Ch>:]CORRection:CONNection
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete
Add
Use "Add" to add a calibration definition to the selected channel. The default calibration
type depends on the channel's measurement mode.
Delete
Select a row in the
table and "Delete" the corresponding calibration defi-
nition.
Cal. Settings
The "Cal. Settings" table displays the calibration definitions of the current channel. The
avaliable columns depend on the available calibration devices.
Figure 5-22: Cal settings for standard calibration units
Figure 5-23: Cal settings for inline calibration systems
Note:
Currently, for each channel either manual calibrations, automatic calibrations, or
inline calibration systems are allowed.
Port <i> ← Cal. Settings
Tick the checkboxes in the "Port <i>" columns to select the ports to be calibrated in the
corresponding calibration.
Note:
Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
●
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
●
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Source Port ← Cal. Settings
Source port for unidirectional n-port calibrations such as "One Path Two Ports" or
"Trans Norm".
Remote command:
First test port in various autocal definition commands such as
Cal softtool