1.
Aim the telescope accurately at
a target at a distance of about
100 m. For distances less than
100 m, make sure to point to
the target precisely. The target
must be positioned within at
least 27°/30 gon above or
beneath the horizontal plane.
☞
The procedure can be
started in face I or II.
☞
The fine aiming must
be performed man-
ually in both faces.
008373_001
V=90°
+ 27°
- 27°
2.
Measure
to measure and to
continue to the next panel.
Motorised instruments change
automatically to the other face.
008372_001
180°
180°
3.
Non-motorised instruments guide to the other face using the
Tele-
scope Positioning
panel.
4.
Measure
to measure the same target in the other face and to calcu-
late the instrument errors.
☞
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and are not averaged with the results from previous runs.
Key
Description
Next
To measure the target.
Description of fields
Field
Option
Description
Number of
measure-
ments
Display only
The number of runs. One run consists of a
measurement in face I and II.
Check & Adjust,
Step 2
272
Settings - TS instrument