22.7.8
Configuring Check & adjust
In
Check & Adjust
,
What do you want to do?
select
Check & adjust set-
tings
and press
Next
.
Key
Description
Next
To accept changes and to continue with the subsequent
panel within the wizard.
Back
To return to the previous panel.
Description of options
Option
Description
Every 2
weeks
,
Once a
month
,
Every 3
months
,
Twice a year
or
Once a year
If one or more adjustment values were determined longer
ago than the time specified with this parameter, then a
reminder message is displayed each time the instrument is
turned on. The reminder helps to redetermine the instrument
errors regularly.
Never
A reminder message to readjust the instrument is never dis-
played. This setting is not recommended.
Next step
Next
to change to the
Report sheet
panel.
22.7.9
Adjusting the Circular Level of the Instrument and Tribrach
4
5b
5b
5b
4
5b
5b
5b
4
008375_001
2
1
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Using the tribrach footscrews, level the instrument with the elec-
tronic level.
3.
Select
Settings
\
TS instrument
\
Level & compensator
to access
the
Level & Compensator
panel.
Access
Check & Adjust
Adjusting the Circular
Level Step-by-Step
276