Key
Description
Edit
To open the panel corresponding to the highlighted field.
Shows the panel that is hidden or displayed.
Next
To accept changes and to continue with the subsequent
panel within the wizard.
Key
Description
Back
To return to the previous panel.
Next
When
Yes, apply user restrictions now
is checked and this
key is pressed, a password can be typed in.
When
No, just finish the wizard
is checked, this key
returns to the
Leica Captivate - Home
.
28.5
Calibration of Internal Sensors
Available for CS20.
Calibrate Disto tilt sensor
is availableif the CS20 is equip-
ped with a DISTO.
Select
Leica Captivate - Home: Settings\System\Calibrate internal sen-
sors
or
Calibrate Disto tilt sensor
.
Calibrate the internal sensors prior to use when the compass functionality is
used.The calibration wizard guides through the calibration process.
Re-calibrate the DISTO tilt sensor if the CS20 has been dropped or heavily
bumped.
The calibration wizard guides through the calibration process.
Steps to check if a re-calibration is required:
1.
Place the CS20 on a flat surface about 20 m away from a wall but
aiming at the wall.
2.
Access a panel where it is possible to view the tilt value: Select
Leica
Captivate - Home: Measure
. Press
Fn
Tools
. Select
Measure hid-
den point
.
3.
The tilt value must be close to 0.
4.
If the tilt differs from 0, then the DISTO calibrate the tilt sensor.
Select the settings to
be available.
Access Restriction
Wizard,
Do you want to apply
user restrictions?
Availability
Access
Sensor Calibration
Disto Tilt Calibration