Key
Description
Back
To return to the previous panel.
Redo
To reject all results and to repeat the complete check and
adjust procedure.
Finish
To accept and store the new determined instrument errors,
which are selected. If the report sheet recording has been
enabled, then the results are written or appended to an
existing report sheet.
Description of metadata
Metadata
Description
New
The new determined and averaged instrument errors.
Old
The old adjustment errors, which are currently valid on the
instrument.
22.7.5
Tilting Axis Adjustment (a)
In
Check & Adjust
,
What do you want to do?
select
Check & adjust the
tilting axis
and press
Next
.
This adjustment procedure determines the following instrument error:
Instrument error
Description
a
Tilting-axis error
Before determining the tilting-axis error, the instrument has to be:
•
levelled up using the electronic level
•
protected from direct sunlight
•
acclimatised to the ambient temperature, approximately 2 minutes per °C
difference compared to the storage place.
•
The horizontal collimation error must be determined before.
Key
Description
Measure
To measure the target.
Access
Description
☞
Check & Adjust,
Step 1
Settings - TS instrument
271