Field
Option
Description
a T-axis
quality (1
σ
)
Display only
The standard deviation of the determined
tilting-axis error. The standard deviation can
be calculated from the second run onwards.
Measure at least two more runs.
Next step
IF
THEN
more runs must be added
select
Add another calibration
loop
and press
Next
.
no more runs must be added
select
Finish the calibration &
store the results
and press
Next
to
accept the measurements and to
access the results panel.
Key
Description
Back
To return to the previous panel.
Finish
To accept and record the new determined tilting-axis error. If
the report recording has been enabled, then the results are
written to or appended to an existing report sheet.
Redo
To reject the result and to repeat the complete check and
adjust procedure.
Description of fields
Field
Option
Description
New
Display only
The new determined and averaged tilting-
axis error.
Old
Display only
The old instrument error, which is valid on
the instrument.
22.7.6
Compensator Adjustment (l, t)
In
Check & Adjust
,
What do you want to do?
select
Check & adjust the
compensator
and press
Next
.
The compensator adjustment procedure determines the following instrument
errors:
Instrument error
Description
l
Compensator longitudinal index error
t
Compensator transversal index error
Before determining the compensator index errors, the instrument has to be:
Check & Adjust,
It is recommended to
repeat the last cali-
bration routine at
least three times
Check & Adjust,
Results
Access
Description
☞
Settings - TS instrument
273