levelled up using the electronic level
•
protected from direct sunlight
•
acclimatised to the ambient temperature, approximately 2 minutes per °C
difference compared to the storage place.
Key
Description
Next
To measure the target.
1.
Measure
to measure the first
face. No target has to be aimed
at.
Motorised instruments change
to the other face and release a
measurement automatically.
008374_001
180°
2.
Non-motorised instruments guide to the other face using the
Tele-
scope Positioning
panel.
3.
Measure
to release the measurement in the other face.
☞
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and are not averaged with the results from previous runs.
Key
Description
Next
To measure the target.
Description of fields
Field
Option
Description
Number of
measure-
ments
Display only
The number of runs. One run consists of a
measurement in face I and II.
l Compo-
nent quality
(1
σ
)
and
t
Component
quality (1
σ
)
Display only
The standard deviations of the determined
adjustment errors. The standard deviations
can be calculated from the second run
onwards.
Measure at least two more runs.
Next step
IF
THEN
more runs must be added
select
Add another calibration
loop
and press
Next
.
Check & Adjust,
Make first tilt meas-
urement in any face
Check & Adjust,
Step 2
Check & Adjust,
It is recommended to
repeat the last cali-
bration routine at
least three times
274
Settings - TS instrument