Next step
IF you want to
THEN
determine the instrument errors
select one of the three available
check and adjust procedures and
refer to the relevant subchapters.
view the current values
select
View the current values
.
Refer to "22.7.7 Viewing the Current
configure
Check & adjust
select
Check & adjust settings
.
Refer to "22.7.8 Configuring Check
adjust the circular level
Refer to "22.7.9 Adjusting the Circu-
lar Level of the Instrument and Tri-
inspect the laser plummet
Refer to "22.7.11 Inspecting the
Laser Plummet of the Instrument".
adjust the tripod
Refer to "22.7.12 Servicing the Tri-
22.7.4
Combined Adjustment (l, t, i, c, ATRplus and Telescope Camera)
In
Check & Adjust
,
What do you want to do?
select
Check & adjust the
compensator, index error, line of sight error & automatic target aiming
or
Check & adjust the compensator, index error, line of sight error, auto-
matic target aiming & telescope camera
and press
Next
.
The combined adjustment procedure determines the following instrument
errors in one process:
Instrument error
Description
l, t
Compensator longitudinal and transversal index
errors
i
Vertical index error, related to the standing axis
c
Horizontal collimation error, also called line of
sight error
ATRplus Hz
ATRplus zero point error for horizontal angle
ATRplus V
ATRplus zero point error for vertical angle
Telescope camera Hz
Telescope camera zero point error for horizontal
angle - option
Telescope camera V
Telescope camera zero point error for vertical
angle - option
Before determining the instrument errors, the instrument has to be:
•
levelled up using the electronic level
•
protected from direct sunlight
•
acclimatised to the ambient temperature, approximately 2 minutes per °C
difference compared to the storage place.
Access
Description
☞
268
Settings - TS instrument