14.4
Connection Using Internal Radio
Description of fields
Field
Option
Description
Link number
Editable field
The assigned channel number.
Set as
Remote
or
Base
☞
The radio modules inside the field
controller and the TS instrument
must be set to opposite settings.
Set the field controller to
Remote
and TS instrument to
Base
.
Next step
Next
and follow the instructions on the panel.
14.5
Connection To Leica Legacy and Third Party Total Stations
Once you begin working with the CS always work on the controller! Do not
touch the total station software, excluding turning the laser pointer, laser
plummet or guide lights on/off for some models.
Refer to "33.7 Connection to Other Total Stations" for supported functions.
Before using any Leica Legacy or third-party total station, ensure that the fol-
lowing values are set on the
TS
:
Instrument
Settings
Leica Legacy total station
1. Total station ppm/scale:
•
Atmospheric ppm = 0
•
Geometric ppm = 0 or scale factor = 1
These settings ensure that the correct
coordinates are calculated at the CS. It is
still possible to apply the relevant atmos-
pheric and geometric ppm/scale factor val-
ues. These values must then be set on the
CS.
2. Communication settings:
•
The communication settings on the TS
must match the default parameters for
that particular instrument type as seen
on the CS.
•
For TPS1000 and TPS1100 instruments:
•
set the communication mode to GSI
•
ensure the TS is in the measurement
panel when trying to connect.
TS Connection Wizard
- Step 2
☞
☞
Settings required
Connections - TS connection wizard
159