IF
THEN
no more runs must be added
select
Finish the calibration &
store the results
and press
Next
to
accept the measurements and to
access the results panel.
Key
Description
Back
To return to the previous panel.
Redo
To reject all results and to repeat the complete check and
adjust procedure.
Finish
To accept and record the new determined instrument errors.
If the report sheet recording has been enabled, then the
results are written and appended to an existing report sheet.
Description of metadata
Metadata
Option
Description
New
Display only
The new determined and averaged instru-
ment errors.
Old
Display only
The old instrument errors, which are currently
valid on the instrument.
22.7.7
Viewing the Current Values
In
Check & Adjust
,
What do you want to do?
select
View the current val-
ues
and press
Next
.
Key
Description
OK
To return to
Check & Adjust
,
What do you want to do?
.
More
To display information about the date of the determination,
the standard deviation of the errors and the temperature
during the determination.
The temperature of the environment around the instrument can differ from the
temperature shown on the panel as it is the internal temperature of the instru-
ment.
Check & Adjust,
Results
Access
Check & Adjust
☞
Settings - TS instrument