The PowerSearch filter can be switched on and switched off in the
Aim &
Search
icon pop-up bubble.
After clicking the
Filter off
bubble icon, the filter is deactivated, but not
deleted. The filter can be switched on again using
Filter on
.
When no filter is learned, the bubble icons
Filter on
and
Filter off
are not
shown.
When pressing the
Filter learn
bubble icon, the instrument performs a Power-
Search scan. The scan order is:
•
400 gon PowerSearch rotation around Hz with a V angle of 100 gon
•
400 gon PowerSearch rotation around Hz with a V angle of 60 gon
•
400 gon PowerSearch rotation around Hz with a V angle of 140 gon
When learning the filter, a previous filter is deleted.
When measuring new points for a setup, the current filter is updated.
The PowerSearch filter is reset after a new setup.
The PowerSearch filter is kept when shutting down and restarting the instru-
ment.
The PowerSearch window can be defined individually. It is specified by absolute
angle values and does not change its position. The PowerSearch window can
be set in the
Target Search, PowerSearch window
page by aiming at two
opposite points of the PowerSearch window. When
Use PowerSearch win-
dow
is checked and a PowerSearch is started, a prism is searched for within
the defined window.
When
Use PowerSearch window
is not checked and the instrument has lost
lock, after the prediction time, the prism is searched for in a dynamic Power-
Search window. This window covers a region at the position after prediction of
horizontal 100 gon by vertical 40 gon.
The PowerSearch routine can be activated clockwise or anticlockwise by using
hotkeys. This action will have no influence on the prism search settings.
33.3
Follow Moving Prisms - Lock
Lock enables instruments equipped with automatic aiming to follow a moving
prism. The automatic aiming sensor is active when Lock is active. When
Target
lock on
is selected in the icon bar, an automatic aiming search is executed.
The instrument locks onto the prism and follows its movements. Automatic
aiming offsets are continuously applied to the angle measurements. When the
instrument loses lock to the prism, a PowerSearch or fine search (auto aiming
search) can be executed depending on the prism search settings.
Lock is unavailable for SmartStation.
PowerSearch window
Dynamic PowerSearch
window
Direction of search
Description