LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
397
5.5.1.1
Pinholes Section Panel
The
Pinholes Section
panel has the following options:
Description
field
Display of the relevant active channel.
Diameter [µm] slider
Position X Vis [µm] slider
Position Y Vis [µm] slider
Position X InVis
[µm] slider
Position Y InVis
[µm] slider
Setting of diameter, X-, Y- and Z-position of the pinhole in relation to
the beam path (Z-position can be set only for PH1) using the slider or
arrow buttons
Status display for setting procedure: green for ready and red for busy.
Stored Pos
button
Pinhole setting is reset to the position last stored.
Adjustment: Adjust
Automatically
button
Automatic pinhole adjustment (not available in some systems).
Fast Adjust Mode
check box:
If this check box is activated, the pinhole adjustment is only performed
in a limited area. Used for readjustment (not available in some
systems).
No further software function can be activated and executed while automatic pinhole adjustment is
running.