ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
272
000000-2071-464
10/2014 V_01
Reuse
With a click on the
Reuse
button, the configuration used to record the selected measurement will be
loaded.
Reload
With
Reload
the current active measurement will
be reopened. If raw data have been stored and
attached to the measurement the
Correlate
window will appear (Fig. 363).
You can redefine the following settings:
−
Automatic dust filter
−
Correlator settings
−
Count rate settings
−
Photon count histogram settings
Only those measurements will be
recalculated with the new settings which
have been selected in the original table. All
other measurements will be opened
unchanged. To select all measurements use
the table context menu opened with a right
mouse click in the table and choose the line
Select all
.
If no raw data have been stored or are
associated with the .fcs file or the
measurement, an error message will be
displayed (Fig. 364). No manipulation of the
data is possible. Click
Ok
to continue
working with the unmodified data set.
Automatic dust filter
If the
Automatic cut
box is checked, you can type in a threshold in % in the
Automatic cut ratio
display box. This sets the threshold intensity for an electronic dust filter. The value (0 … 100) will define
the threshold in percentage. A default value of
10 %
can be selected by pressing the button on the right.
Fig. 363
Correlate window
Fig. 364
Correlate Error window
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......