ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
218
000000-2071-464
10/2014 V_01
You can choose between symmetric rotation
(1 exponential term) and asymmetric rotation
(double exponential term) by selecting / deselecting
the
Asymmetric
check box (Fig. 285).
If the
Asymmetric
box is checked, you can define
relative amplitudes and rotational frequencies
(note, these parameters are no fit values and have
to be defined by the user).
Relative amplitudes are defined by selecting the
excitation polarization (
linear polarized
or
unpolarized
) and the emission detection
(
parallel
,
perpendicular
or
all
), by pressing the
corresponding tabs, which will result in the
corresponding values displayed in the
c1
and
c2
display boxes. Alternatively, if
others
is activated,
you can type in user defined values.
Rotational frequencies can be selected by pressing
the
Default
tab to load the default values. The
corresponding values will be shown in the
r1
and
r2
display boxes. By pressing the
other
tab, user
defined values can be entered.
The description box gives information on the
rotational diffusion terms.
Fig. 285
Rotation Settings panel,
Asymmetric rotation checked
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......