QUASAR detector settings
Use the two sliders to define the spectral detection range of the QUASAR detector.
Resolution
: Step size can be 3.0, 4.4, 8.9, 17.8, 26.7 or 35.6 nm.
Online Fingerprinting Mode
(to be selected in the
Imaging Setup
tool).
The use of this function permits the selection of
reference spectra together with the excitation
settings, allowing an immediate display of the
unmixed images while scanning. Chose
Online
Fingerprinting
in the
Imaging Setup
tool.
You can change the settings using the function
elements which have been described for the
Lambda Mode
plus the following:
RS1 ... 8
Menu for selecting a display color and reference
spectra (reference spectra derive from earlier
experiments via Mean of ROI functionality or ACE
tool). The unmixed results are displayed during
scanning and Lambda Stack data will neither be
displayed nor stored.
Fig. 116
Imaging Setup tool window;
Online Fingerprinting mode