LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
199
A.4
Amplitude A
1
)
0
(
−
=
=
G
N
A
γ
(5a)
where
γ
is the geometric factor accounting for the point spread function (PSF) and N the mean number of
particles.
In the FCS software
γ
can be a fit or a predefined fixed value. In case
γ
is a fit value N must be fixed in the
fit procedure. N is normally a fit parameter.
Please note that
γ
takes different values for different fitting models depending on the assumed intensity
distribution of the point spread function (PSF):
γ
C
=1.000 (cylindrical),
γ
2DG
=0.500 (2-D Gaussian),
γ
3DG
=0.350 (3-D Gaussian),
γ
GL
=0.076 (Gaussian-Lorentzian).
γ
can also be calibrated, if a known concentration c of a dye is measured. In this case N can be fixed and
γ
fitted. The obtained number can be entered as the calibrated fixed number. N can be calculated from
equation
A
L
V
N
c
⋅
=
(5c)
with V being the confocal volume and L
A
= 6.023 x 10
23
mol
-1
the Avogadro number.
The volume V is calculated from equation
z
r
V
ω
ω
π
⋅
⋅
=
2
2
3
(5d)
with
ω
z
axial focus radius and
ω
r
the lateral focus radius. The radii themselves have to be determined by a
calibration measurement using a dye with a high quantum yield and a known diffusion coefficient D from
the fitted diffusion time
t
d
and the structural parameter S employing a free diffusion model with triplet
state.
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......