LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
235
•
Continue moving upwards slowly. The line
position will reach its maximum (Fig. 311).
•
Having slowly moved over a short distance the
line position will decline again to a minimum
and than rise again to a second maximum
(Fig. 312), which corresponds to the upper
cover slip surface.
•
Store the reached position by using the
Store
button.
•
Now you can position the detection volume into
the sample in a well-defined way:
If you use one of the C-Apochromat water
immersion objectives, position the detection
volume 150 - 200 µm deep into the sample to
get rid of disturbing interface effects. The
easiest way of doing this is by clicking the
+200 µm
button.
Fig. 311
Line diagram at lower cover slip
surface
Fig. 312
Line diagram at upper cover slip
surface