LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
231
5.3.8
Navigation – Focus
The function of the
Focus
tool allows you to
position the focus in Z direction.
•
Press the
Focus
tool tab to open the
Focus
tool
(Fig. 304).
Press anywhere outside the window to close the
tool box.
The following functions are available:
Zero Position
display field (only available if
Show
all
is checked): Displays the current z-position. L
indicates load, H home position. The arrows will
move the focus continuously u and down as long
as pressed.
Step Size [µm]
allows you to enter the steps the
focus will move.
Z-Position
box displays the actual z-position. The
arrows will move the focus with a step size defined
in
Step Size [µm]
.
Stop
button will interrupt an active z-movement.
Work
button: Moves the stage / nosepiece back to
the Work position. This is the position last set
before the
Load
button was pressed.
Load
button: Lowers the stage / nosepiece to a default position make it easier for you to change the
sample carrier (or objective).
Zero Position Manually
button will set the current z-position to zero.
If a z Piezo-stage insert is available the
Piezo
field will be displayed. The
Piezo Position [
µ
m]
box
displays the actual z-position of the Piezo-stage insert. The arrows will move the focus with a step size
defined in
Step Size [µm]
. Pressing the
Levelling
button will move the middle position of the z Piezo-
stage to the current z-position of the microscope stand's nosepiece.
Mark position
button will store the current z-position.
Reuse #.#
button will move to the z-position that was defined by
Mark position
. The z position number
#.#
will appear only, when
Mark position
has been pressed and indicates the z coordinate of the
present position.
The Z-position stored by activation of the
Mark position
button is a relative parameter and is only
valid during the current session. Accordingly, this position can be approached in a defined way only
during this session via
Reuse
.
+200 µm
button: Moves the Z drive for +200 µm per mouse click upwards.
-200 µm
button: Moves the Z drive for -200 µm per mouse click downwards.
If the sample is just a solution without structure, one cannot "focus" the instrument by looking at sample
features. Cover slip reflection has to be used instead to find the glass- / solution interface. The position of
the glass surfaces is detected most conveniently via line scanning.
Fig. 304
Focus tool
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......