ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
384
000000-2071-464
10/2014 V_01
5.4.21
Processing – ICS
Licensing
Instrument
standard
optional
LSM
ELYRA
Lightsheet Z.1
Tool accepts exclusively input acquired with LSM 710 and LSM 780. See full
description of the tool set and section on
RICS (Raster Scanning Image Correlation
Spectroscopy)
for more details on requirements to be met.
•
Activate
ICS
in the Processing tool by clicking
on the arrow.
The
ICS
menu provides the following sub-
processing functions:
Detrend
,
ICS correlation,
Map filter, NandB Analysis and Remove
structures
5.4.21.1
Detrend
•
The Detrend function is a high pass recursive
filter.
•
Click
Detrend
in the
ICS
tool. The
Method
Parameters
menu will show the input options
for this function.
•
Select the desired image in the image container, then click
Select
to define this image as the
Input
image.
The resuling output image using the time constant parameter is shown in the
Preview
tool and the
Preview View tab in the image container.
Input images are time series images.
The
Time constant
used to detrend the image is the number of images it takes for the recursive filter to
decay to 1/e. Use a smaller number for detrending faster moving larger objects like vesicles and a higher
number for detrending bleaching effect or slower cell movements.
Fig. 540
Processing – ICS
Summary of Contents for LSM 880