LSM 880
Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
357
Using the channel alignment tool for drift corection
The following procedure can be used to remove lateral
drift in an inage:
•
Duplicate the time series. Select the time series in the
processing tool
Copy
under the submenu
Duplication
(Fig. 507). Press
Apply
to obtain a copy
from your time series.
•
Generate reference image. Using the Processing tool
Copy
select the duplicate time series in the submenu
Subset
. Using
Time
select your reference image, for
example the first image or the center image, by
choosing for the
Start
and
End
value the same
number (Fig. 508). Press
Apply
to extract that image
from the time series.
•
Generate reference time series. In the Proceesing tool
Modify
choose under
Modify mode
the entry
Make
time series
. Select the exctracted image. Enter under
Number of times
and
Time intervals
the numbers
from your original time series (Fig. 509) and press
Apply
. The result will be a time series with the same
image at each time point.
Fig. 507
Methods Parameter of the
Copy Duplicat Processing tool
Fig. 508
Methods Parameter of the
Copy Subset Processing tool
Fig. 509
Methods Parameter with
Make time series selected in
Modify mode
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......