Left Tool Area and Hardware Control Tools
LSM 880
254
000000-2071-464
10/2014 V_01
5.3.17
Count Rate
•
Click the
Count rate
button to open the
Count
rate
window (Fig. 339).
a)
Display
Select one of the display modes
Count rate
,
Correlation
or
CPM
(Counts / molecule) by
pressing the corresponding button. The arrow
buttons of the scale tab allow to adjust the
measuring range.
•
Click the
Count rate button in the
Display
line to display the count rate per second (CR)
for each activated physical channel (Fig. 339).
This value is constantly updated and calculated
each 0.5 ms, if
Averaging
is switched
Off
. If
system overload occurs, an error message will
be displayed and the APDs switched off in order
to avoid damage (Fig. 340). Press
Ok
to
continue and laser light intensity or
concentration of your sample.
•
Click the
Correlation
button to display the
correlation amplitude (G(0)) of each activated
physical and Software channel (Fig. 341). This
amplitude is constantly updated and calculated
each 0.5 ms if
Averaging
is switched
Off
. The
expanded display for each channel can be
closed by the
Channel 1
,
Channel 2
and Cross
Correlation arrows.
Fig. 339
Count rate window with count rate
display
Fig. 340
Overload display
Fig. 341
Count rate window with correlation
display