LSM 880
Startup and Shutdown of the System
ZEISS
10/2014 V_01
000000-2071-464
25
3.2.2
Switching System Power Off
After having exited the ZEN software, you may turn off the whole system with the following
Shutdown
procedure.
•
Make sure you have saved all your image data.
•
Switch of all the lasers which might still be running.
•
Close the ZEN software window. This terminates the ZEN software.
•
Shut down WINDOWS.
•
About 20 seconds after WINDOWS is shut down your computer turns off.
Please bear in mind that a cooling phase of at least 5 minutes is required between switching off
of the Argon laser and switching off of the entire system via the power remote switch or the
Power Supply switch.
5 minutes after computer shutdown
set the power remote switches
System/PC
and
Components
or the
Main
switch on the System Electronic Rack to position
OFF
.
This puts your LSM 880 including the computer off power.