ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
342
000000-2071-464
10/2014 V_01
The
PSF sampling
pull down menu allows you to
interpolate the result to different degrees. Since
the resulting PSF is an average of many PSFs
corresponding to the localized beads and every
bead is localized with the sub-pixel accuracy, one
can obtain after averaging a PSF with higher
resolution.
You can average the PSF by
1
(pixel size of the
resulting PSF is the same or
1
times as in the
original image),
3
(pixel size of the resulting PSF is
3
times smaller as in the original image),
5
(pixel
size of the resulting PSF is
5
times smaller as in the
original image) or
7
(pixel size of the resulting PSF
is
7
times smaller as in the original image)
times (Fig. 471).
The
3D PALM
box will be automatically checked, if
the image data was acquired with the 3D-PALM
slider inserted. In this case the double lobe
structure of each bead will be aligned to the right
lobe.
In the
Background
drop down menu you can
decide the way background is determined
(Fig. 472).
You have three options:
−
Local
: Background is computed from the
periphery area of the mask defined in
Peak
Mask Size
.
−
Histogramm
: Background is calculkated
from the histogram as at 10% cut off value.
−
ROI
: Backgrund is compited from ROIs in the
image (
drawn
with the Graphics tool). Note,
this option is only available if ROIs have been
defined in the image.
There are different filters that can be set by corresponding sliders or spin input boxes to select sub-
resolution beads.
Fig. 470
Experimental PSF – manual mode
Fig. 471
PSF sampling drop down menu
Fig. 472
Background drop down menu
Summary of Contents for LSM 880
Page 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Page 650: ......
Page 678: ......
Page 687: ......
Page 688: ......