Image View Controls – Table Tab
The
Table
tab (Fig. 365) chosen from the
Option tool area
displays several options:
−
In the
Average
display window you can select, how the data are averaged. You can either select
Fit results
, in which case the single values of the parameters are averaged, or
Correlation results
,
where the single data points of each correlation functions are averaged and an average correlation
function is calculated, from the drop down menu. You can check, from what measurements
(
Repetitions
,
Kinetics
,
Positions
,
Correlation
) you want to built the average from by checking
the corresponding boxes.
−
In the
Order
display window you can choose, how different measurements are grouped together.
The hierarchy of grouping of categories is determined by their listed sequence from the drop down
menu.
−
If the
Show standard deviation
box is checked, all parameters will be displayed with their values
and the calculated standard deviation from different measurements.
Only in the latter case the displayed average correlation function can be meaningfully fitted.
The
Displayed columns
selection window allows you to change the appearance of the table with regard
to columns to be displayed. All parameters with their boxes checked will be visible in the table. The order
of the displayed columns can be adjusted using drag and drop in the table.
Each row contains a check box on the left-hand side. Deactivation of these check boxes will exclude the
relevant rows from all subsequent evaluation procedures like averaging. These settings will be stored by
saving the data. Immediate reactivation is possible via the check box, if required. The scroll bar allows
viewing all content of the display box.
Fig. 365