Left Tool Area and Hardware Control Tools
LSM 880
316
000000-2071-464
10/2014 V_01
5.4.9.2
Ratiometric Dyes
•
Fura-2, Indo-, SNARF, Cameleon, Ratiometric Pericam, Phluorin.
•
Display fluorescence ratio R over time
•
Display fluorescence ratio R corrected for background/autofluorescence over time
•
Calculate absolute ion concentrations (pixel by pixel) via titration calibration (known ion concentrations
applied to the cells – in situ – or in solutions – in vitro or equation calibration where possible [Fura-2,
Indo-, SNARF]
•
Calculation of R eliminates artifacts and uncertainties caused by
−
inhomogeneous dye distribution
−
photobleaching
−
may be applied with moving cells.
Ratiometric Dyes - Online Ratio
R(t1) = F1(t1) / F2(t1), R(t2) = F1(t2) / F2(t2) .
Fig. 418
Ratiometric Dyes – online ratio