Left Tool Area and Hardware Control Tools
ZEISS
10/2014 V_01
000000-2071-464
281
5.3.21.8
Fit
•
Click on the
Fit
tab to activate the
Fit
display mode (Fig. 375).
In general this menu allows you to work with already measured and probably also fitted correlation
curves. The intention is to have the possibility of performing other fit using different fitting parameters if
the automatic fit did not yield optimal results.
The panel is divided in Diagrams, Result Table, Model Field and Fit action buttons.
(1)
Diagrams
Fig. 375
Fit diagram panel for correlation analysis
For fitting to a correlation model the upper graph (Fig. 375) shows the
Correlation G(
t
) Fit diagram
(correlation plotted versus the correlation time) with the overlaid fit graph. It also displays the fit range
defined by the red (start value) and the blue (end value) bars. The lower
Fit deviation
curve depicts the
fit residuals. If an experiment is loaded for which no fit has been conducted yet, only the measured
Correlation G(
t
)
curve is shown and no overlaid fit curve.
The Fit deviation
diagram will be empty.