11
-
7
Operating Temperature : 15 to 35
°
C
Operating Humidity
: 30 to 85% (non-condensing)
Storage Temperature : –5 to 50
°
C
Storage Humidity
: 15% to 90% (non-condensing)
Altitude
: –1,000 to 10,000ft (–300 to 3000m)
Interfering Gases for Anesthetic Agents
The monitor will report small changes in agent concentrations when anesthetic agents and other
gases are used.
Expected agent changes are provided here for the purpose of comparison.
For Gas Mixtures of 5% CO
2
Agents
Agent Volume
Change of CO
2
N
2
O 60%
+0.2
Halothane 4%
+0.1
Enflurane 5%
+0.1
Isoflurane 5%
+0.1
Sevoflurane 5% +0.2
Desflurane* 5% 0.0
Helium 50%
0.0
Ethanol 1%
+0.1
Isopropanol 1% 0.0
Acetone 1%
+0.1
Methane 1%
0.0
Metered dose inhaler propellant
1%
0.0
80% Xenon
Not intended for use with Xenon
N/A
For Gas Mixtures of 30% O2
Agents
Agent Volume
Change of CO
2
N
2
O 60%
–0.1
Halothane 4%
–0.2
Enflurane 5%
–0.2
Isoflurane 5%
–0.2
Sevoflurane 5% –0.2
Desflurane* 5% –0.2
Helium 50%
–0.3
Ethanol 1%
–0.2
Isopropanol 1% +0.5
Acetone 1%
+0.1
Methane 1%
–0.1
Metered dose inhaler propellant
1%
+0.2
80% Xenon
Not intended for use with Xenon
N/A
11
Specifi
cation /
Performan
ce
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...