Alarm Setup
To maintain the alarm setting even after the power is turned OFF or after a discharge procedure, store
the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the Discharge
Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
●
CO
2
Concentration Alarm
EtCO
2
Alarm
The alarm setup should be performed for each measurement
unit. (mmHg/kPa/%). The upper/lower limit can be set in
1mmHg/0.1kPa/0.1% increments.
Lower
Limit
Range: 1 to 98mmHg/0.1 to 13.1kPa/0.1 to 13.1%.
Turns OFF if 1mmHg/0.1kPa/0.1% or below is set.
Upper
Limit
Range: 3 to 115mmHg/0.3 to 15.0kPa/0.3 to 15.0%.
Turns OFF if 115mmHg/15.0kPa/15.0% or above is set.
Auto
Automatically sets the upper limit to
+10mmHg/1.3kPa/1.3%, and lower limit to
–10mmHg/1.3kPa/1.3% from the current value.
In-CO
2
Alarm
The alarm setup should be performed for each measurement
unit. (mmHg/kPa/%). The upper limit can be set in
1mmHg/0.1kPa/0.1% increments.
Lower
Limit
(No Setting)
Upper
Limit
Range: 1 to 24mmHg/0.1 to 3.0kPa/0.1 to 3.0%.
Turns OFF if 24mmHg/3.0kPa/3.0% or above is set.
Auto
Automatically sets the upper limit to
3mmHg/0.4kPa/0.4%.
6
Multiga
s Data
(MGU-7
01/7
02/Poet IQ, MGU-80
1P/8
02/803
)
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...