Remote Control Setup
This section explains the setup procedure to use the optional remote control unit, CF-700.
About the Remote Control Unit
There are 11 keys on the remote control unit of which 8 keys are user keys.
Home Key
Monitor Selection Dial
Alarm Silence Key
Function Key
(User Key)
Check Key
Monitor Selection Dial
The monitor ID to control can be selected. The remote control monitor ID should be preprogrammed
on each bedside monitor. Maximum of 8 monitors can be controlled with one remote control unit.
Check Key
Displays the remote control monitor ID on the bedside monitor. Also, pressing one of the function keys
(ex. NIBP Start/Stop key) will display a confirmation message on the monitor which will ask you to
press this Check key.
Alarm Silence Key
This key does not function on the DS-7000.
Home Key
This key is the same as the Home key on the bedside monitor.
Function Key (F1 to F8)
The function for each key can be assigned on the bedside monitor.
Key Default
Function
F1 ECG1
Size
Switches the ECG1 size each time the key is pressed.
×
1/4
→
×
1/2
→
×
1
→
×
2
→
×
4
→
×
1/4
F2 ECG1
Lead
Switches the ECG1 lead each time the key is pressed.
3-electrode:
Ι
→
ΙΙ
→
ΙΙΙ
→
Ι
4-electrode:
Ι
→
ΙΙ
→
ΙΙΙ
→
aVR
→
aVL
→
aVF
→
Ι
5-electrode:
Ι
→
ΙΙ
→
ΙΙΙ
→
aVR
→
aVL
→
aVF
→
V
→
Ι
F3 NIBP
Start/Stop
Starts/stops the NIBP measurement.
Pressing this key will display a confirmation message on the monitor.
When the “Check” key is pressed, the measurement will start.
To cancel the process, press the “Home” key.
Pressing the “NIBP Start/Stop” key during measurement will stop the
measurement.
F4
Record Start/Stop
Starts/stops the manual recording.
F5
Night Mode ON/OFF
Sets ON/OFF of the Night Mode.
F6
Table
Displays the data in tabular format.
F7
Trend
Displays the trend data.
F8
Switch Display
Switches the display mode.
9
-
18
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...