Mean BP Value Display
The mean BP value can be set to be displayed inside the NIBP numeric data box.
1
Select ON/OFF for “Mean BP Display”.
ON will display the mean BP value.
OFF will not display the mean BP value.
C A U T I O N
If the mean BP value is not displayed, the alarm for mean BP will not
generate. The mean BP value will not be displayed for the table data either.
●
Pulse Rate Display
The pulse rate measured during NIBP measurement can be displayed. This pulse rate is only for
display. It will not generate an alarm, or be displayed for the table function.
1
Select ON or OFF for “PR Display”.
ON will display the pulse rate.
OFF will not display the pulse rate.
Pulse Rate