NIBP Alarm
1
Press the NIBP Alarm key.
Set ON/OFF of the NIBP alarm, upper and lower alarm limits of systolic (S), diastolic (D), mean (M)
NIBP.
The upper and lower limit can be set in 5mmHg / 0.5kPa
increment.
Key
Item
Description
Individual Alarm
Selecting ON will generate a NIBP alarm.
Selecting OFF will not generate a NIBP alarm.
▲
Current Value
Indicates the current value.
Select from S (systolic BP), D (diastolic BP), or M (mean
BP).
Lower Alarm Limit
Sets the lower alarm limit.
Setting the value below the range will turn OFF the alarm.
(See below for NIBP alarm range.)
Upper Alarm Limit
Sets the upper limit.
Setting the value above the range will turn OFF the alarm.
(See below for NIBP alarm range.)
Automatic Setup
Automatically sets the upper limit to +5.0kPa
from the current value, and the lower limit to
–20mmHg/–2.5kPa from the current value.
【
NIBP Alarm Range
】
Systolic
Lower Limit
60 to 245mmHg (Adult)
40 to 175mmHg (Child)
30 to 115mmHg (Neonate)
8.0 to 32.5kPa (Adult)
5.0 to 23.5kPa (Child)
4.0 to 15.5kPa (Neonate)
Upper Limit
65 to 250mmHg (Adult)
45 to 180mmHg (Child)
35 to 120mmHg (Neonate)
8.5 to 33.0kPa (Adult)
5.5 to 24.0kPa (Child)
4.5 to 16.0kPa (Neonate)
Mean
Lower Limit
45 to 230mmHg (Adult)
30 to 155mmHg (Child)
15 to 95mmHg (Neonate)
6.0 to 30.5kPa (Adult)
4.0 to 20.5kPa (Child)
2.0 to 12.5kPa (Neonate)
Upper Limit
50 to 235mmHg (Adult)
35 to 160mmHg (Child)
20 to 100mmHg (Neonate)
6.5 to 31.0kPa (Adult)
4.5 to 21.0kPa (Child)
2.5 to 13.0kPa (Neonate)
Diastolic
Lower Limit
40 to 195mmHg (Adult)
20 to 145mmHg (Child)
10 to 85mmHg (Neonate)
5.0 to 25.5kPa (Adult)
2.5 to 19.5kPa (Child)
1.0 to 11.5kPa (Neonate)
Upper Limit
45 to 200mmHg (Adult)
25 to 150mmHg (Child)
15 to 90mmHg (Neonate)
5.5 to 26.0kPa (Adult)
3.0 to 20.0kPa (Child)
1.5 to 12.0kPa (Neonate)
To maintain the alarm setting even after the power is turned OFF or after a discharge procedure, store
the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the Discharge
Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
6
Non
-In
va
sive Blood Pre
ssu
re
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...